Search options

Search
Search by author
Between Dates
to
Volume and Number
&
Spemd Logo

Revista Portuguesa de Estomatologia, Medicina Dentária e Cirurgia Maxilofacial

SPEMD | 2016 | 57 (4) | Page(s) 197-206


Original research

Analysis of the stability of 3Y-TZP zirconia abutments after thermocycling and mechanical loading

Análise da estabilidade de pilares de zircónia 3Y-TZP após carga cíclica e termociclagem


a Faculty of Dental Medicine, University of Porto, Porto, Portugal.
b Faculty of Engineering, University of Porto, Porto, Portugal.
c Center for MicroElectroMechanical Systems-MEMS, Department of Mechanical Engineering, Minho University, Guimarães, Portugal
d INEGI, University of Porto, Porto, Portugal
e Faculty of Dental Medicine, University of Porto, Porto, Portugal

  Show More



Volume - 57
Issue - 4
Original research
Pages - 197-206
Go to Volume


Received on 13/05/2016
Accepted on 27/06/2016
Available Online on 22/10/2016


Objectives: Although implant abutments made of yttria-stabilized tetragonal zirconia polycrystal (3Y-TZP) show great predictability, some issues related to its aging require further study. The objective of this in vitro study is to assess the stability of 3Y-TZP zirconia implant abutments with two different implant connections when subjected to aging simulation through thermocycling and mechanical loading (TCML). Methods: Ten 3Y-TZP zirconia abutments were selected and equally divided into two groups (n = 5): CEZr, abutments for externally hexed implants; and CIZr, abutments for implants with an internal conical connection. The samples were subjected to thermocycling (5000 cycles; 5–55 °C) and mechanical loading (1.2 × 106 cycles; 88.8 N; 4 Hz). Before and after the aging procedures, X-ray diffraction (XRD) analysis was conducted to observe tetragonalmonoclinic (t-m) phase transformation, and topographic surface analysis was performed by 3D profilometry, and data were analyzed using Mann-Witney test (p < 0.05). Results: XRD measurements revealed no monoclinic phase in any of the abutments after aging. The comparative analysis regarding roughness (using the Sa parameter) at the abutments’ seating platforms (using 3D profilometry) revealed a slight increase in roughness in both connections after TCML. Statistically significant differences (test U = 57.0, p = 0.161 > 0.05), before and after TCML, and between implants connections (test U = 57.0, p = 0.053 > 0.05) were not detected. Conclusions: After a 5-year simulation of its clinical use, the analyzed 3Y-TZP zirconia abutments did not show signs of aging. The connection’s geometry does not interfere in aging.


Objetivos: Os pilares de zircónia tetragonal policristalina (Y-TZP) apresentam grande previsibilidade na sua utilização clínica. No entanto, existem ainda alguns aspetos relacionados com o envelhecimento que necessitam de ser estudados. O presente trabalho pretendeu avaliar a estabilidade de pilares de zircónia 3Y-TZP de 2 conexões implantares diferentes, quando submetidos a simulação de envelhecimento (TCML). Métodos: Foram selecionados 10 pilares em zircónia 3Y-TZP, que foram divididos em 2 grupos (n= 5): CEZr, pilares para implantes de conexão de hexágono externo; e CIZr, pilares para implantes de conexão interna cónica-lobular. As amostras foram submetidas a termociclagem (5000 ciclos; 5-55 °C) e carga cíclica (1,2 × 106 ciclos; 88,8 N; 4 Hz). Antes e após TCML, os pilares foram sujeitos a análise por difração de raios X para determinação de alteração de fase cristalográfica de tetragonal para monoclínica (t-m) e a interferometria ótica sobre a superfície de assentamento das 2 conexões para a medição da sua topografia por perfilometria 3D. Análise estatística: Mann-Whitney test (p < 0,05). Resultados: Após TCML não foi detetada fase monoclínica em nenhum dos pilares. A análise comparativa da rugosidade (com utilização do parâmetro Sa) sobre as plataformas protéticas dos pilares revelou ligeiro aumento da Sa em ambas as conexões após TCML. Não foram observadas diferenças estatisticamente significativas nos valores de Sa (test U = 57,0, p = 0,161 > 0,05), antes e após TCML, e entre conexões dos implantes (test U = 57,0, p = 0,053 > 0,05). Conclusões: Após uma simulação de 5 anos de utilização clínica, os pilares de zircónia 3YTZP analisados não apresentaram sinais de envelhecimento. A geometria das conexões não interferiu no envelhecimento.


Supplementary Content


  Download PDF